E1.Initial reporter facility name: (b)(6).The device was returned for evaluation.A visual and tactile examination identified multiple kinks along the length of the hypotube shaft, no kinks or damages to the shaft polymer extrusion, and no balloon damage.The device was received with the product stylet/mandrel.The stylet/mandrel was not inserted through the wire lumen and an examination of the tip section of the device revealed the tip was detached from the balloon catheter and was positioned on the stylet/mandrel.A microscopic examination of the tip break noted that the break is consistent with excessive tensile forces being applied to the tip, resulting in the break.Microscopic examination of the distal extrusion, balloon, and blades identified no damage.
|
Reportable based on device analysis completed on (b)(6) 2023.It was reported that the mandrel could not be removed.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During preparation, it was noted that the mandrel used to protect the tip during shipment could not be removed.The procedure was completed with a different device.There is no patient injury reported.However, device investigations revealed that the tip was detached from the balloon catheter.
|