Catalog Number 320559 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.D.4.Medical device expiration date: unknown.H.4.Device manufacture date: unknown.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle would not attach.The following was translated from japanese to english: this is a complaint about needle does not stick into cartridge and cannot be installed.When the customer tried to attach it to the pen, they could not do so.The needle did not stick into the cartridge.
|
|
Manufacturer Narrative
|
H.6.Investigation summary: samples were received and an investigation was performed.Embecta was able to duplicate or confirm the indicated issue.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
|
Event Description
|
It was reported that the bd nano¿ 2nd gen pen needle would not attach.The following was translated from japanese to english: this is a complaint about needle does not stick into cartridge and cannot be installed.When the customer tried to attach it to the pen, they could not do so.The needle did not stick into the cartridge.
|
|