• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problems Failure to Form Staple (2579); Failure to Fire (2610); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic cholecystectomy, the trigger could not be grasped when firing.An unexpected sound was heard when firing, and the twisted clip came out.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4).Date sent: 12/19/2023 d4: batch # unk attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: (b)(6) 2024 d4: batch # (b)(4).Investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 10 conforming clips. upon testing, the jaws open and close without any difficulties.In addition, the device locked out as intended.  although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torquing may result in clip malformation.Do not insert the clip applier through a trocar if a clip is present in the jaws.This may result in clip malformation, dislodged clips, or damage to the instrument.If a clip is present in the jaws, fully squeeze the trigger against the handle, then fully release the trigger to release the clip from the jaws before inserting the device through the trocar.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18361371
MDR Text Key331079867
Report Number3005075853-2023-09644
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberA9DF8Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-