Catalog Number 306572 |
Device Problem
Short Fill (1575)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that bd syringe 10ml saline xs did not have the designated amount of saline.The following information was provided by the initial reporter: the customer states that, "visited patient for disconnection.Syringe removed out of packaging for flushing purposes.On removal, noticed syringe had only 6ml instead of stated 10ml on packaging".
|
|
Manufacturer Narrative
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion a supplemental report will be filed.
|
|
Event Description
|
No additional information.
|
|
Manufacturer Narrative
|
A device history record review was completed for provided material number 306572 and lot number 3089846.The review did not reveal any possible non-conformances during the production process that could have contributed to this reported incident.To aid in the investigation of this issue, four (4) picture samples were received for evaluation by our quality team.The picture samples confirmed the defect of underfilled syringe.Although we were able to confirm the defect, an exact cause for the underfilled syringe could not be determined.If the physical sample becomes available, we would appreciate the opportunity to review it.This is the first report received for this type of issue on material (b)(4) and lot 3089846.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
|
|
Manufacturer Narrative
|
Correction made: we became aware on (b)(6) 2024 that the awareness date for this complaint was (b)(6) 2023.
|
|
Search Alerts/Recalls
|