• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 54F; PROTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 54F; PROTHESIS, HIPS Back to Search Results
Model Number N/A
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It was reported the cup cannot be removed from the shell inserter.The surgeon removed the cup and implanted a new shell causing a thirty (30) minute delay.No known impact or consequence to the patient.
 
Manufacturer Narrative
(b)(4).G2: foreign: china.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-03020.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d4, d9, g3, g6, h2, h3, h4, h6.Visual examination of the returned product identified there is bio-debris inside of the threads.There is damage to the threads of the device due to the damage the device could not be checked with a gauge.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.Definitive root cause cannot be determined.Complaint confirmed based on the damage to the returned product.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 54F
Type of Device
PROTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18363245
MDR Text Key330976735
Report Number0001825034-2023-03018
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304524231
UDI-Public(01)00880304524231(17)330808(10)7560173
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010000664
Device Lot Number7560173
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Age75 YR
Patient SexMale
-
-