MAUDE Adverse Event Report: LDR MEDICAL MOBI-C IMPLANT "M" STANDARD 15X15 H5 US; MOBI-C CERVICAL DISC PROTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MEDICAL MOBI-C IMPLANT "M" STANDARD 15X15 H5 US; MOBI-C CERVICAL DISC PROTHESIS Back to Search Results
Catalog Number MB3555
Device Problems Fracture (1260); Malposition of Device (2616)
Patient Problems Pain (1994); Numbness (2415)
Event Type  Injury  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that a revision surgery was performed for a mobi-c construct that was causing pain and numbness in the patient.During the revision surgery it was discovered that the left anterior corner of the mobile core had fractured, allowing the core to shift alignment and collapsing the posterior portion of the c3-c4 construct.The mobi-c was removed and replaced with an acdf.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBI-C IMPLANT "M" STANDARD 15X15 H5 US
Type of Device
MOBI-C CERVICAL DISC PROTHESIS
Manufacturer (Section D)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR  10300
Manufacturer (Section G)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR   10300
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18364910
MDR Text Key331000918
Report Number3004788213-2023-00143
Device Sequence Number1
Product Code MJO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110009
Exemption Number5645646
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberMB3555
Device Lot Number5324606
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient EthnicityNon Hispanic
-
-