Internal reference number: (b)(4).H10: this complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from literature sources that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.Kyle alpaugh, yu-fen chiu, jason p.Zlotnicki, ilya bendich, alejandro gonzalez della valle, mathias p.G.Bostrom, elizabeth b.Gausden, femoral component undersizing and alignment are risk factors for early periprosthetic femur fracture, the journal of arthroplasty, volume 37, issue 7, supplement, 2022, pages s604-s610, issn 0883-5403, https://doi.Org/10.1016/j.Arth.2022.03.015.
|
It was reported that on literature review "femoral component undersizing and alignment are risk factors for early periprosthetic femur fracture", after undergoing a tha with a synergy cementless stem, fifteen (15) patients sustained a periprosthetic femur fracture.Patients¿ outcome and how it was treated is unknown.No further information is available.
|
It was reported that on literature review "femoral component undersizing and alignment are risk factors for early periprosthetic femur fracture", after undergoing a tha with a synergy cementless stem, fifteen (15) patients sustained a periprosthetic femur fracture.Patients required revision surgery.Patients¿ outcome is unknown.No further information is available.
|