EDWARDS LIFESCIENCES EDWARDS COMMANDER DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9610TF23U
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
As reported by an edwards lifesciences affiliate in france, regarding a 23mm sapien 3 ultra valve in aortic position by transfemoral approach.During the procedure, at the end of deployment with nominal volume, the 23mm commander delivery system balloon burst.The valve was correctly implanted at correct position and fully expanded.The delivery system was retrieved through the esheath with no difficulties.The patient did not suffer any injury and was noted as to have no impact due to the event.
 
Manufacturer Narrative
Investigation is underway.H3 other text : not returned.
 
Manufacturer Narrative
The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.A previous investigation into this type of event is captured in an edwards lifesciences technical summary and applies to this complaint.Additional assessment of the failure mode is not required at this time.The device was returned for evaluation.The returned device was visually examined, and the following was observed: balloon longitudinal and radial burst at 40 cm from nose tip.Due to the nature of the complaint (balloon burst), no functional testing was able to be performed.The complaint was confirmed through visual examination.The inflation balloon single wall thickness was measured along the edges of the burst location.All measurements taken of the balloon single wall thickness met the specification.Per the technical summary, the ifu, current risk mitigations include design and manufacturing controls, and training manuals have been reviewed, no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported event was confirmed based on evaluation of the returned device.Available information suggests that patient factors (calcification) likely contributed to the event as the provided information mentioned presence of calcification in patients native valve.The presence of calcification can create a challenging anatomy for balloon inflation.While the balloons are sufficiently designed and tested to ensure the burst pressure is at or above the rated burst pressure, calcified nodules can compromise the structure of the balloon wall via following mechanisms such as puncture, local overstretching, open cell impingement, or stress concentration.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Since no edwards product defects or labeling deficiencies were identified, no corrective or preventative action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS COMMANDER DELIVERY SYSTEM
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506384
MDR Report Key18365853
MDR Text Key331318313
Report Number2015691-2023-18592
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9610TF23U
Device Lot Number65181532
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-