Model Number CLV-S400 |
Device Problems
No Device Output (1435); Connection Problem (2900); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned as part of the asset return process, in addition to the reportable findings in b5, the following findings were discovered; the front panel is crushed and the top cover was deformed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
|
Event Description
|
A user facility returned the olympus asset, xenon light source, to the olympus service center.Upon inspection and testing of the returned device, it was observed that due to deformation of the scope socket, the endoscope could not be identified, the base did not hold the light guide cable, and the high function was not available.This report is being submitted for the event found during evaluation.There was no patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eight (8) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the light cable is not recognized due to the deformed scope socket.Olympus will continue to monitor field performance for this device.
|
|