• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER SKIN GRAFT MESHER; EXPANDER, SURGICAL, SKIN GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER SKIN GRAFT MESHER; EXPANDER, SURGICAL, SKIN GRAFT Back to Search Results
Model Number N/A
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.G2 foreign: australia.Review of the most recent repair record determined the technician confirmed the wrench cannot slide into mesher.The roller and ratchet gear were damaged and were replaced which resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the wrench cannot slide onto device.The event occurred at an unknown time.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER SKIN GRAFT MESHER
Type of Device
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18369361
MDR Text Key331216438
Report Number0001526350-2023-01669
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375390
UDI-Public(01)00889024375390(11)190404(10)64371003
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00770100000
Device Lot Number64371003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-