Model Number N/A |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.G2 foreign: australia.Review of the most recent repair record determined the technician confirmed the wrench cannot slide into mesher.The roller and ratchet gear were damaged and were replaced which resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that the wrench cannot slide onto device.The event occurred at an unknown time.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
|
|