The customer reported a false negative alinity i total b-hcg result for one patient sample while running on the alinity i processing module.The following data was provided (= 5.00 miu/ml is negative, >/= 25.00 miu/ml is positive): the initial b-hcg result in the morning was < 2.4 miu/ml (negative).The doctor called and questioned the patient¿s negative result.The customer reran the sample on another alinity instrument, and the result was 2000 miu/ml (positive).All the qc was within range.There was no impact to patient management reported.
|
After further evaluation, the suspect medical device was changed from alinity i total b-hcg reagent kit, 07p51-21, longford to alinity i processing module, 03r65-01, irving ia/cc.Mfr report number 3016438761-2024-00059 has been submitted and all further information will be documented under that mfr report number.H3 other text : after further evaluation, the suspect medical device was changed from alinity i total b-hcg reagent kit, 07p51-21, longford to alinity i processing module, 03r65-01, irving ia/cc.Mfr report number 3016438761-2024-00059 has been submitted and all further information will be documented under that mfr report number.
|
The customer reported a false negative alinity i total b-hcg result for one patient sample while running on the alinity i processing module.The following data was provided (= 5.00 miu/ml is negative, >/= 25.00 miu/ml is positive): the initial b-hcg result in the morning was < 2.4 miu/ml (negative).The doctor called and questioned the patient¿s negative result.The customer reran the sample on another alinity instrument, and the result was 2000 miu/ml (positive).All the qc was within range.There was no impact to patient management reported.
|