Model Number 331/5661
Device Problems Crack (1135); Material Separation (1562); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2023
Event Type  malfunction  
Manufacturer Narrative
D10 concomitant product: 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); 331/5661, 331/5661 extend swivel catheter mt x25 (lot#22g0267fax); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, a few minutes after connecting to the machine, the ventilator had an alarm because the inner interface wall was broken and the interface of the swivel catheter cracked.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.Nine devices and two photos were available for evaluation.Visual inspection found a cracked 15mm connector.It was reported that a few minutes after connecting to the machine, the ventilator had an alarm because the inner interface wall was broken and the interface of the swivel catheter cracked.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAR
Type of Device
CONNECTOR, AIRWAY (EXTENSION)
Manufacturer (Section D)
MALLINCKRODT DAR SRL
via giacomo bove 2/4/6/8
mirandola 41037
IT  41037
Manufacturer (Section G)
MALLINCKRODT DAR SRL
via giacomo bove 2/4/6/8
mirandola 41037
IT   41037
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18373717
MDR Text Key331188727
Report Number2936999-2023-01187
Device Sequence Number1
Product Code BZA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K942392
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number331/5661
Device Catalogue Number331/5661
Device Lot Number22G0267FAX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/27/2023
Date Device Manufactured07/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-