RESPIRONICS, INC. EVERFLO; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1020007
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2023
Event Type  malfunction  
Event Description
The manufacturer received information alleging an everflo device had a defective compressor.A device was returned to the manufacturer.During the evaluation of the device at the manufacturer, the device was visually inspected and found the sieves clogged, springs rusted, capacitor is defective and pca is defective.The parts were replaced, and the device was repaired.The inlet filter was repaired due to preventative maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18377600
MDR Text Key331209036
Report Number2518422-2023-37282
Device Sequence Number1
Product Code CAW
UDI-Device Identifier00606959032569
UDI-Public00606959032569
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K061261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1020007
Device Catalogue Number1020007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-