• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems No Display/Image (1183); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2023
Event Type  malfunction  
Event Description
It was reported that there was a malfunction resulting in the loss of display.There was no report of patient involvement.
 
Manufacturer Narrative
Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Block d4 unique identifier: (b)(4).Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.H3 other text : device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
Service will be performed by hospital employee or contractor.A ge healthcare service representative recommended the high performance display unit board to be replaced to resolve the issue.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The display fan was replaced to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key18377985
MDR Text Key331330437
Report Number2112667-2023-06834
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-