• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Catalog Number 9-PFO-3025
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pericardial Effusion (3271)
Event Date 11/28/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2023, a 30-25mmamplatzer talisman pfo occluder was chosen for a patent foramen ovale (pfo) closure using a 9f amplatzer talisman delivery system.During procedure, heparin was administered and the wire across the pfo was difficult and multiple attempts were made with non abbott wires and a non-abbott sheath.The difficulty was due to anatomy of the patient.The sheath was appeared to be in good location but the device did not sit properly due to patient anatomy.The device was not device released from the delivery system and another attempt was made to cross septum.After crossing, an angiogram was performed and a contrast was noted around the heart.A pericardial effusion was noted and monitored no intervention performed.The procedure was aborted.The patient was reported stable.
 
Manufacturer Narrative
An event of pericardial effusion was reported.There was difficulty crossing the septum or positioning the device due patients anatomy.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the received information the cause of reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18378231
MDR Text Key331181540
Report Number2135147-2023-05611
Device Sequence Number1
Product Code MLV
UDI-Device Identifier05415067033321
UDI-Public05415067033321
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number9-PFO-3025
Device Lot Number8670485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-