BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10015
Device Problems Connection Problem (2900); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a cori assisted tka surgery, a "system time out" error was displayed.After this notification the real intelligence robotic drill did not function any more.They were able to take the burr out the real intelligence robotic drill but could not unlock the ri robotic drill attachment nor the real intelligence robotic drill tracker.They also tried to unlock them in the admin but did not work.The surgery was completed, after a non-significant delay, changing to manual procedure.No injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI ROBOTIC DRILL ATTACHMENT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18378320
MDR Text Key331378889
Report Number3010266064-2023-00240
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB10015
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
RI CORI(PN:ROB10024, SN(B)(6)).; RI ROB DRLL TRCKR(PN:ROB10014).; RI ROB DRLL(PN:ROB10013, SN(B)(6)).
-
-