BOSTON SCIENTIFIC CORPORATION EMBOLD FIBERED DETACHABLE COIL SYSTEM; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
Model Number 83910 |
Device Problems
Stretched (1601); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/13/2023 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on 05dec2023.It was reported that the coil would not advance in the microcatheter and stretched.A 6x20 embold fibered detachable coil system was selected for use in a transsplenic gastric variceal embolization procedure.The microcatheter was flushed; however, the coil would not advance with the microcatheter.The whole system was pulled out.The physician regained access and deployed other embold coils successfully.There were no patient complications.However, device analysis revealed the coil was separated from the distal coupler.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the delivery sheath, delivery wire and a severely damaged, stretched embold 6x20 coil were returned to boston scientific for analysis.Visual and microscopic analysis was conducted.The perforations were intact.The delivery wire was inside of the delivery sheath with the stretched coil.The distal and proximal couplers were attached to the delivery wire via the pull wire.No damage on the delivery sheath and delivery wire was noticed.The couplers showed slight bends.The coil was severely stretched and separated from the distal coupler.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|
|
|