MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA HYSTEROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA HYSTEROVIDEOSCOPE Back to Search Results
Model Number HYF-V
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the evaluation found additional non-reportable findings were as follows: leak at biopsy port, bending section cover (a-rubber) was peeling on both sides, the image was intermittently flickering when moving the bending section, smashed/dent on the insertion tube, scratches on light guide tube (lg tube), scratches on video cable, peeling labels on scope connector, dents and scratches on video connector.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the hysterovideoscope biopsy port spun, making it difficult to remove the stop cock when attached.The issue was found during reprocessing.There was no delay to the procedure and the procedure was completed.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, and the information provided, the cause of the loose biopsy port could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA HYSTEROVIDEOSCOPE
Type of Device
HYSTEROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18381019
MDR Text Key331384601
Report Number3002808148-2023-14635
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04953170340185
UDI-Public04953170340185
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221557
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHYF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-