ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2023
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported during the intraocular lens (iol) implant procedure noticed a scratch on the iol.The iol was explanted and replaced with another lens.The iol exchange went smoothly.There was no patient harm.Additional information has been received that the iol was implanted and removed on the same day.According to nurse, the scratch could have been caused by the injector (plunger) and cartridge.
 
Event Description
There are two medical device reports associated with this patient.This report is associated with the left eye.
 
Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The file indicates the use of a qualified cartridge and viscoelastic.The file also indicates the use of a non-qualified handpiece.The root cause is most likely a failure to follow the ifu (instructions for use).The account used an non-qualified lens/cartridge/handpiece combination.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18384140
MDR Text Key331226650
Report Number1119421-2023-02202
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF VISION IOL.; ASICO ROYAL INJECTOR.; BSS STERILE SOLUTION.; PROVISC.
Patient SexFemale
-
-