SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/085
Device Problem Separation Problem (4043)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2023
Event Type  malfunction  
Event Description
It was reported that air escaped from the cuff.When the cuff was before using it, the cuff deflated.During the patient's intubation, the cuff seemed to be deflated, making it easier for the tube to fall out.There was no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
D9 - date returned to mfg; 1/2/2024.H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One used decontaminated device sample was received for investigation without its original packaging.Under visual inspection the sample appeared to be in good condition.An inflation test was performed on the received sample.It was confirmed that after twelve (12) hours the cuff was still fully inflated.The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action taken and no trend of similar customer complaints was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT SUCTIONAID
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL CZECH REPUBLIC A. S
olomoucká 306
hranice 753 0 1
EZ  753 01
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18384542
MDR Text Key331487108
Report Number3011237704-2023-00014
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076165
UDI-Public15019517076165
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/085
Device Catalogue Number101/860/085CZ
Device Lot Number4310532
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-