• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RAD-G PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RAD-G PULSE OXIMETER Back to Search Results
Model Number 27977
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2023
Event Type  malfunction  
Event Description
The customer reported the rad-g "turns on and off on its own." there was no patient impact or consequence reported.
 
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned rad-g was evaluated.The device intermittently powered on then powered off shortly after powering on.This was due to shorting inside the on/off power button which created an electrical short across the button and resulted in the button being activated even when not physically pressed.
 
Event Description
The customer reported the rad-g "turns on and off on its own." there was no patient impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-G PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
tahereh sedighi
15750 alton pkwy
irvine, CA 92618
9492977862
MDR Report Key18384877
MDR Text Key331492461
Report Number3019388613-2023-00280
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997000666
UDI-Public00843997000666
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201770
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27977
Device Catalogue Number9849
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-