Additional information: d9, g3, h6.Complaint is not confirmed.One unpackaged ar-6480 dualwave arthroscopy fluid management system was returned for investigation.The returned device was assembled with a new ar-6411/ar-6421 and ar-6430 arthroscopy pump tubing and was tested and evaluated under normal use conditions.The pump was powered on, and no error messages and no audible alarms were triggered.A clamping test was performed as defined in the user guide (dfu-0212 rev 1 ¿ section 5.2) to simulate an overpressure failure on the pump and to verify if an error message and/or audible alarm would be triggered.The results of the clamp test indicate that the pump triggered an alarm, and the rollers did stop moving.This behavior is expected.Pressure fault test, when the pressure was artificially lowered by removing the inlet from the fluid source, an alarm was triggered, and the device stopped.This behavior is expected.No problem found.Functional testing with the pressure calibrator found no pump pressure issues.The pump was supplied with 100mmhg and 200mmhg of pressure and displayed the correct reading for each.No problem found.The pump was tested with the shaver console, and the device worked as required.No problem found.The outflow was tested and did not find issues with the suction and lavage system or connection of the cassette.No problem found.Visual evaluation did not find any issues with the device.The review of complaint records for serial number (b)(6) shows that the device has no previous complaints.The original warranty seal was present and completed.The manufacturing date of the device is 2021-11-20.No problem found.Refer to investigation photos.Complaint is not confirmed.
|