• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the hl20 displayed the error message: "runaway" intermittently during routine check.No harm to any person has been reported.A getinge field service technician (fst) was onsite for an investigation and repair on 2023-12-19.The fst was not able to reproduce the reported error message: "runaway".The fst checked and reset all connections of the pump.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.No exact root cause could be determined, because the failure could not be reproduced.With reference to the hl20 risk management file the most probable root cause could be determined as follows: (un)intended change of pump speed by e.G.: deactivated interventions / override; by knob; by display setting; slave mode (controlled by master).The device in question was manufactured on 2018-07-06.The review of the non-conformities during the period of 2018-07-06 to 2023-12-22 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.Based on the results the reported error message: "runaway" could not be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the hl20 displayed the error message: "runaway" intermittently.No harm to any person has been reported.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18389592
MDR Text Key331290012
Report Number8010762-2023-00643
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701028580
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2018
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-