• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM Back to Search Results
Model Number MID-C 125
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2023
Event Type  malfunction  
Manufacturer Narrative
Production process analysis: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.User (surgeon and patient) information analysis: patient #(b)(6) index procedure performed on (b)(6) 2022.On (b)(6) 2023 apifix was notifed that patient #(b)(6) underwent revision surgery on that day due to t6 screw pull-out.During the revision the polyaxial screws were replaced with bigger screws (7.0x35) and moved up one level (t4/t5 to t5/t6).A new mid-c and extender of the same size and type were replaced.The device is not being returned to the manufacturer for analysis and will remain with the patient.Patient requested to keep the implant.Clinical affairs reviewed images from the original surgery and from the pre-op (b)(6) 2023 and could not determine if the kite angle collapsed pulling the screw with it or vice versa.Risk assessment: reoperation events are a known risk that was assessed and recorded by the product risk management file.The risk of screw pullout has been assessed and found to be acceptable the current screw pullout rate due to any reason is in line with the rate reported in the literature for this type of complication as described in the company's cer (clinical evaluation report).The risk has been quantified, characterized, and documented as acceptable within a full risk assessment.The event of screw pull-out (misplaced screws in pedicle/nut loosening from the pedicle screw & subsequent surgical interventions) is addressed in the ifu as potential risks associated with the mid-c system.Apifix is closing this complaint, but will continue to monitor this 'failure mode'; complaint trending will continue to monitor per post marketing surveillance procedure.If any further relevant information is identified, the complaint file will be reopened and a supplemental medwatch will be filed.
 
Event Description
On (b)(6) 2023 apifix was notifed that patient #(b)(6) underwent revision surgery on that day due to screw pull-out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 125
Type of Device
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel street
kochav yokneam bldg
yokneam elit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key18390426
MDR Text Key331583894
Report Number3013461531-2023-00060
Device Sequence Number1
Product Code QGP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H170001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Report Date 12/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMID-C 125
Device Catalogue NumberMUS-125-050
Device Lot NumberAF-01-069-21
Date Manufacturer Received11/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
Patient SexFemale
-
-