As noted in the b.5 response, ronin responded promptly to conduct a thorough investigation.At that time, it also requested information from the surgeon with respect to her use of the device.Based upon its investigation and the surgeon's response to ronin's use questions, ronin can only conclude that its device was not the cause of the reported burn.
|
This incident is the subject of medical device report mw5144689, which was submitted to fda on august 14, 2023 by dr.(b)(6).A copy of this mdr was provided to ronin by fda on september 12, 2023.On may 13, 2023, ronin surgical received an email communication from dr.(b)(6) informing ronin of a surgery performed several days prior during which the patient experienced a burn.A ronin x6 headlight was reportedly used during the procedure.It was further reported that the minimum distance the device was used at was greater than 12 inches from the patient.On or about may 15, 2023, during a phone call between ronin and (b)(6) hospital for children, a mutual decision was made by the parties that the hospital would return the headlight used in the surgery to ronin for evaluation.Ronin received the headlight used in the surgery, its holster, and its battery on or about may 25, 2023.Ronin conducted a manual inspection of the components and found no evidence of physical damage or mechanical abnormality.Ronin also conducted a functional test and found no evidence of abnormal behavior or performance.Thermocouple ("heat rise") testing was also conducted and confirmed that the alleged subject device was performing within specifications and was not defective; ronin therefore ruled out the headlight as a cause of the reported issue.At the hospital's request, ronin provided a report detailing its findings to the hospital on or about june 5, 2023.As of ronin's last communication with (b)(6) hospital for children's vp of quality on or about july 10, 2023, no additional information was provided by the hospital, and the results of the hospital's investigation were that they were not able to determine the cause of the burn.Disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a ronin surgical product malfunctioned, is defective or has caused serious injury.
|