• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION BSM-1733A; VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION BSM-1733A; VITAL SIGNS MONITOR Back to Search Results
Model Number BSM-1733A
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2023
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the respiratory rate (rr) value is changing constantly on this device.According to the customer, the value will drop to 0 and then it goes back up.The customer has tried changing out the leads and used the same leads on a different 1700 unit and that unit worked normally.The customer will send in the unit to be exchanged.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the respiratory rate (rr) value is changing constantly on this device.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BSM-1733A
Type of Device
VITAL SIGNS MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18392919
MDR Text Key331310154
Report Number8030229-2023-03954
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921111833
UDI-Public4931921111833
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBSM-1733A
Device Catalogue NumberBSM-1733A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-