The device was sent to the philips authorized repair facility (rft) for bench for repair.Results of functional testing indicate that speaker produced audible sound.Based on the information available and the testing conducted, the cause of the reported problem was not replicated.A good faith effort was then performed, and information was received confirming that the speaker was producing sound at the time of the event.Although the speaker was confirmed to be functioning per specification during testing, the speaker has been replaced per current process.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.Based on the information received, indicating the speaker did actually produce sound at the time of the event, this is no longer considered a reportable event.
|