Model Number 865350 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that speaker had no sound and beep using speaker certification tool.The cause of the reported problem was a defective speaker.The reported problem was confirmed.He speaker was replaced.The device was operational after repairs were completed and the device was returned to the customer.
|
|
Event Description
|
During evaluation at bench repair, it was identified that the device had no audio.The device was not in use on a patient at the time of the event.There was no adverse event reported.
|
|
Manufacturer Narrative
|
This complaint was reported in error.Sound was present at bench repair and no report is needed.
|
|