The device was returned and evaluated.In addition to the malfunction documented in b5, the additional evaluation findings are as follows: due to a pinhole on the ch-tube, water tightness was lost; due to the wear of the angle wire, the bending angle in the up direction did not meet the standard value; due to the wear of the forceps lever, the up/down knob could not be locked securely; the control unit had corrosion due to water leakage; damage or deformation due to physical stress; the ultrasonic transducer had corrosion and the connecting tube and switch had scratches.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility. .
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed foreign material in the channel tube and distal end of the device could not be identified and a definitive root cause of the issue could not be determined, however, due to the observed leak in the channel tube, it is possible that proper reprocessing could not be performed.The issue may be detected/prevented by following the instructions for use below: instruction manual evis lucera ultrasonic gastrovideoscope olympus gf type uct260 chapter 6 application and conditions of cleaning, disinfection, and sterilization.Chapter 7 cleaning, disinfection, and sterilization procedures.Reprocessing survey info: the device was cleaned, disinfected, and sterilized before being sent to olympus.No delay in the start of pre-cleaning.No abnormalities in the accessories used for reprocessing.The air/water nozzle was wiped/brushed with clean lint free cloths, brushes, or sponges.Olympus will continue to monitor field performance for this device.
|