ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Electrical Power Problem (2925); Activation Problem (4042)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2023 |
Event Type
malfunction
|
Event Description
|
It was reported outside of surgery that when the lever is depressed to start the motor, nothing happens.After evaluation of the returned device, it was determine that the dermatome's speed was inconsistent.There was no harm, injury, or delay as there was no patient involvement.Due diligence is complete.No additional information is available.No adverse event reported.
|
|
Manufacturer Narrative
|
This event is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the switch works in the opposite direction and the motor speed was unstable; however, the repair has not been completed because the repair site is waiting for quotation approval.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.G2, country event occurred in: south africa.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|