Catalog Number ROB10013 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a cori-assisted tka, while cutting the distal femur, they received "robotic drill deactivated" error from the real intelligence robotic drill.They tried to press continue to see if it would work and it let attempt to continue to burr.However, got the error again.This happened a few times.In-between receiving the errors, the burr seemed like it was attempting to cut but it could not.In addition, after the case was complete, it was ran the troubled drill through a kpc test and it failed the "maximum speed" portion which was not surprising because while holding the drill during that part of the test, the burr was having trouble turning and i could feel the motor struggling.Surgery was performed after a delay greater than 30 min, with a back-up device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H3, h6: the real intelligence robotic drill, part number rob10013, serial number (b)(6) , used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with drill motor or drill motor shaft failure.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|
Search Alerts/Recalls
|
|