• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNK STEM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNK STEM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Corroded (1131)
Event Date 11/29/2023
Event Type  Injury  
Event Description
It was reported the patient underwent an initial hip procedure.Subsequently, the patient was revised approximately 10 years later due to elevated ion levels.An mri showed a collection of fluid and debris.The trunnion of the stem showed signs of corrosion and discoloration.It was reported that no further information is available.
 
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2023-00315.D10: cat# 00630505636, lot# 62387796, liner standard 3.5 mm offset 36 mm i.D.For use with 56 mm o.D.Shell.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further information at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04) stem.The stem was not returned and remains implanted.Pictures provided show dark discoloration and bio-debris to the trunnion.No further evaluation can be made from the provided pictures.Part and lot identification are necessary for review of device history records, neither were provided for the stem.Medical records were not provided.Image assessed but not sent to mmi.Trunnion corrosion is not visible on xray and explant/intraop images within the subforms provide sufficient evidence of findings.Sending the image would not enhance the investigation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK STEM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18399072
MDR Text Key331401606
Report Number0001822565-2023-03744
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNK STEM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight94 KG
-
-