Catalog Number MCS20
Device Problems Difficult to Open or Close (2921); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 12/27/2023.B3: unknown, assumed first day of month that complaint was reported.D4: batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during an unknown procedure, the jaws did not open, and the clip could not be fed into the jaw, although the clips were remained.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Manufacturer Narrative
(b)(4).Date sent: 1/31/2024.D4: batch # a9d964.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcs20 device was returned with no damage to the external components.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was disassembled to evaluate the condition of the internal components and the feedbar was found damaged, causing the firing issue.Six(6) clips were found inside the clip track.Although no conclusion could be reached on what caused the damage to the feed bar, it is possible that the device was clipped over a hard object resulting in higher force to fire, subsequently damaging the mechanism.Please note that the instructions for use contain the following caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18399116
MDR Text Key331402122
Report Number3005075853-2023-09872
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036002499
UDI-Public10705036002499
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCS20
Device Lot NumberA9D964
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-