ABBOTT MEDICAL FLEXNAV DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Catalog Number FNAV-DS-LG |
Device Problems
Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that on (b)(6) 2023, a 27mm portico valve was successfully implanted, using a large flexnav delivery system.During removal of the flexnav from the descending aorta, the nose cone was noted to disengage, requiring femoral surgical access to retrieve it.The patient is reported to be stable.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
|
|
Manufacturer Narrative
|
An event of nose cone being disengaged during removal of delivery system from descending aorta and requiring femoral surgical access to retrieve it was reported.One image from field appeared to show detached inner shaft that appeared damaged.It is unlikely for the nosecone to detach from the member unless the member itself was kinked during retraction.A returned device inspection, to rule out any device-related causes, could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
|
Search Alerts/Recalls
|
|
|