Model Number 72204878S
Device Problems Material Deformation (2976); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2023
Event Type  malfunction  
Event Description
According to the reporter, the hysteroscope's shaft had cracked and sharp edges.There was no patient involvement.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the outer tube had dents, was bent, compressed, had distal tip damage, there were raised fibers, the welded seam was cracked and there were broken lenses in the optical system with less than 50% of the lenses defective.Functional testing found that the root cause for these damages were customer related issues.It was reported that the hysteroscope's shaft had cracked and sharp edges.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN MANSFIELD
15 hampshire street
mansfield MA 02048
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18407411
MDR Text Key331709708
Report Number1282497-2023-00076
Device Sequence Number1
Product Code HIH
UDI-Device Identifier10884521767744
UDI-Public10884521767744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180496
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72204878S
Device Catalogue Number72204878S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-