Model Number A-SP3-03K
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Laceration(s) (1946)
Event Date 10/09/2023
Event Type  Injury  
Manufacturer Narrative
"to date, spatz fgia inc.Has not received the product for evaluation, therefore no analysis or testing has been done.A review of the device labeling notes the following: · proper positioning of the spatz3 adjustable balloon system® within the stomach is necessary to allow proper inflation.Lodging of the balloon in the esophageal opening during inflation may cause injury and/or esophageal rupture.The physiological response of the patient to the presence of the spatz3 adjustable balloon system® may vary depending upon the patient's general condition and the level and type of activity.The complication section refers to: · esophageal obstruction.Once the balloon has been inflated in the stomach, the balloon could be pushed back into the esophagus.If this occurs, surgery or endoscopic removal could be required.Injury to the digestive tract during placement of the balloon in an improper location such as in the esophagus or duodenum.This could cause bleeding or even perforation, which could require a surgical correction for control.· injury to teeth, tissue in the oral cavity or throat and upper esophageal sphincter.".
 
Event Description
By introducing the intragastric balloon without complications, filling the balloon after confirming the position in the stomach.At the end of filling, there was a spontaneous rupture of the connector's internal hose.The balloon was deflated with a needle and the loop was removed.We observed lacerations in the mucosa in the teg, requiring clipping to close the lacerations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPATZ3 ADJUSTABLE BALLOON SYSTEM
Type of Device
GASTRIC BALLOON
Manufacturer (Section D)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer (Section G)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer Contact
pnina polishuk
1801 s perimeter rd, ste 130
fort lauderdale, FL 33309
MDR Report Key18408595
MDR Text Key331482215
Report Number3012638928-2023-02956
Device Sequence Number1
Product Code LTI
UDI-Device Identifier860005178810
UDI-Public(01)860005178810(11)2023-04-16(17)2025-04-16(10)230416
Combination Product (y/n)N
PMA/PMN Number
P190012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-SP3-03K
Device Lot Number230416
Was Device Available for Evaluation? No
Date Manufacturer Received11/28/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening; Hospitalization;
-
-