• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SFX,5.5,TI, LAT, SIZE A3; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SFX,5.5,TI, LAT, SIZE A3; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 189401303
Device Problems Mechanical Problem (1384); Material Twisted/Bent (2981)
Patient Problem Insufficient Information (4580)
Event Date 12/07/2023
Event Type  malfunction  
Event Description
Device report from synthes reports an event in korea, south as follows: it was reported on (b)(6) 2023, that in the process of tightening the socket nut, wear occurred, preventing the next step from being carried out.It is unknown if the surgery was delayed due to the reported event.Is unknow if the procedure was successfully completed.This report is for one (1) sfx,5.5,ti, lat, size a3.This is report 1 of 2 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional product codes: osh,kwp,mni,kwq and mnh.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the set screw of the sfx,5.5,ti, lat, size a3 was found stripped.The will not tension allegation can attribute to the stripped condition.A dimensional inspection was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The overall complaint was confirmed as the observed condition of the sfx,5.5,ti, lat, size a3 [189401303/om12034] wouldt contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Drawing/specifications reviewed dwg-1894-01-301/305 rev f current & manufactured.Dimensional inspection: n/a.H4, h6: a review of the receiving inspection (ri) for sfx,5.5,ti, lat, size a3, was conducted identifying that lot number om12034 was released in one batch.Batch1: lot units were released on 31 oct 2022 with no discrepancies.Supplier : (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The surgery was successfully completed.Patient status/outcome: n/a.No broken fragments were retained in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX,5.5,TI, LAT, SIZE A3
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 02400
SZ   02400
3035526892
MDR Report Key18411493
MDR Text Key331705947
Report Number1526439-2023-02643
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034168036
UDI-Public(01)10705034168036
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number189401303
Device Lot NumberOM12034
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SFX,5.5,TI, LAT, SIZE A2
-
-