Model Number CF-HQ190L |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/13/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the colonovideoscope air/water channel was blocked and the lens was a little opaque.The issue was found during an unspecified procedure.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation, and the evaluation found that weak air/water flow was detected.After the nozzle was removed, the air/water flow was normal.The surface dirt was found on the objective lens.The device evaluation also found the control knob had minor play.The distal end had scratches.The light guide lens had tiny chips.The bending section was cracked.There was a trace of water in the scope connector. the investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material was unable to be identified and deviation from instruction for use (ifu) are unknown.Furthermore, physical damage of the nozzle was not confirmed.Therefore, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to insufficient or inadequate reprocessing.The event can be detected and prevented by following the ifu section: detection methods are written in ifu: gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection.Prevention measures are written in ifu: gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|