• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that during preventative maintenance by getinge field service engineer, the cardiosave intra-aortic balloon pump (iabp) unit has damaged fiber optic sensor extension.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
A getinge field service engineer evaluated the unit and discovered that the fiber optic sensor extension cable was damaged while completing the pm.- obtained customer approval and completed billable repair.To fix the issue , fse replaced fiber optic cable assembly.Full measurement details are available on pm so# (b)(4).Device passed all functional and safety tests according to factory specifications.Device was returned to customer.The defective components were received for further investigation.Please refer to the root cause evaluation field for details.The failure analysis and testing dept.Received part number 0012-00-1562 rev.E, with a reported unit failure of a broken fiber optic connector.The fat performed a visual inspection and found the part to be broken where the fiber optic cable plugs in.Please see attachment.Verified the reported issue but no root cause for damage determined.Retaining the part in the failure analysis and testing department per procedure number (b)(4) rev.Ap.The non-conformances with the returned components were confirmed.However, the root cause or the most probable root cause is impossible to be defined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18415805
MDR Text Key331582587
Report Number2249723-2023-05539
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-