DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventative maintenance by getinge field service engineer, the cardiosave intra-aortic balloon pump (iabp) unit has damaged fiber optic sensor extension.There was no patient involvement.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
A getinge field service engineer evaluated the unit and discovered that the fiber optic sensor extension cable was damaged while completing the pm.- obtained customer approval and completed billable repair.To fix the issue , fse replaced fiber optic cable assembly.Full measurement details are available on pm so# (b)(4).Device passed all functional and safety tests according to factory specifications.Device was returned to customer.The defective components were received for further investigation.Please refer to the root cause evaluation field for details.The failure analysis and testing dept.Received part number 0012-00-1562 rev.E, with a reported unit failure of a broken fiber optic connector.The fat performed a visual inspection and found the part to be broken where the fiber optic cable plugs in.Please see attachment.Verified the reported issue but no root cause for damage determined.Retaining the part in the failure analysis and testing department per procedure number (b)(4) rev.Ap.The non-conformances with the returned components were confirmed.However, the root cause or the most probable root cause is impossible to be defined.
|
|
Search Alerts/Recalls
|
|
|