Catalog Number 28-B2-33-100U
Device Problems Failure to Align (2522); Device Misassembled During Manufacturing /Shipping (2912); Manufacturing, Packaging or Shipping Problem (2975); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Event Description
"upon fluoroscopic examination of device before entering the patient it was determined that the device was loaded upside down with the contra limb loaded in the posterior position vs anterior.When physician rotated the system to the patient's left the contra limb max and min flouro markers moved to the patient right vs left.The physician remedied the situation by rotating the device to the patients right to orient the contra limb to the patients left as desired.Physician was relying on the mb flush valve to gauge orientation.Further conversations need to be had with the customer to review how we load our grafts into the delivery system and techniques for contra gate orientation.Customer mentioned that this would be unacceptable on any future fenestrated technology that we would launch." patient outcome - "no clinical sequelae as the graft was safely deployed with the desired gate orientation.".
 
Manufacturer Narrative
Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREO ABDOMINAL STENT-GRAFT SYSTEM
Type of Device
STENT, ENDOVASCULAR GRAFT, AORTIC
Manufacturer (Section D)
BOLTON MEDICAL, INC.
799 international parkway
sunrise FL 33325
Manufacturer Contact
megan indeglia
799 international parkway
sunrise, FL 33325
9548389699
MDR Report Key18416085
MDR Text Key331641657
Report Number2247858-2023-00331
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P190015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number28-B2-33-100U
Device Lot Number2309290216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/28/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-