Model Number CARDIOHELP.I
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the planetary gear shaft of the emergency drive was corroded.The failure occurred during preventive maintenance.A getinge field service technician (fst) was sent for investigation and repair.The planetary gear was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The root cause of the defective planetary gear is corrosion on the gear shaft.It is stated in the instruction for use of the cardiohelp device (chapter 5.6.1 "check before every application") that the emergency drive must be checked before use.The review of the non-conformities has been performed on 2023-12-22 for the period of 2017-07-24 to 2023-12-15.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2017-07-24.Based on the results the reported failure "emergency drive shaft corroded" could be confirmed.This complaint was found in the database of customer complaints for the cardiohelp device as a single event (timeframe from 2022-12-15 till 2023-12-15).The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the planetary gear shaft of the cardiohelp emergency drive has corrosion.No harm to any person has been reported.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18416173
MDR Text Key331590601
Report Number8010762-2023-00646
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP.I
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-