• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device was found to be microbiologically contaminated during weekly test of the water.There is no report of patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6) belgium.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
Throught follow up livanova learned the device is cleaned regularly.The hospital does not use reusable blankets, the water quality monitoring is applied and the h2o2 checked every day as prescribed by the ifu, the device is in use every day, the h2o2 is checked daily but not in the weekend, h2o2 is added when the water in the tanks is changed (each 7 days), a disposable pall-aquasafe water filter with an 0.2 m membrane (or equivalent performance filter) is used for tap water, prior to initial operation and prior to storing the heater-cooler the surfaces and water circuits are disinfected and device surfaces are disinfected once a week, the 3t aerosol collection set is replaced after the allowed use period.The device is placed outside the operation theatre during use.As per the above, the device is not cleaned after every operations and h2o2 is checked every day but not in the weekend.This is not in accordance with ifu.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
It cannot be excluded that highlighted deviation from ifu's may have contributed to reported contamination.Livanova has implemented a strategy to progressively decrease the probability of bacteria growth in the hc device by applying multiple measures implemented over the past few years through dedicated capa and field action.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key18417462
MDR Text Key331595779
Report Number9611109-2023-00663
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public(01)04033817901099(11)170529
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/16/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-