• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC LIBRE SENSOR FREESTYLE; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC LIBRE SENSOR FREESTYLE; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71536-01
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2023
Event Type  malfunction  
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The dhrs (device history review) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBRE SENSOR FREESTYLE
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18417768
MDR Text Key331578013
Report Number2954323-2023-55523
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number71536-01
Was Device Available for Evaluation? No
Date Manufacturer Received12/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-