Model Number 865241
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Event Description
The customer reported speaker malfunction inop.The device was in use at time of event, there was no adverse event reported.
 
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Manufacturer Narrative
Philips received a complaint on the intellivue mx700 patient monitor indicating that there was a speaker malfunction.A good faith effort (gfe) was performed to clarify if the customer was alleging sound issues with the speaker or only a speaker malfunction inop on the monitor, but no additional information was provided.The device was sent to philips bench repair.A philips bench repair technician (brt) evaluated the device and was unable to confirm the issue.No speaker malfunction error was appearing on screen.The brt confirmed that sound was emitting from the device properly at different volumes.There was no trouble with the device during testing.Based on the information available and the testing conducted philips was unable to replicate the reported problem.The reported problem was not confirmed.The device was confirmed to be operating per specifications and no failure was identified.The brt replaced the speaker for preventative measure.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Event Description
The customer reported a speaker malfunction inop.It is unknown if sound was still coming from the device.The device was in use at time of event, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX700 PATIENT MONITOR
Type of Device
INTELLIVUE MX700 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18418422
MDR Text Key331599429
Report Number9610816-2023-00692
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838029088
UDI-Public00884838029088
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865241
Device Catalogue Number865241
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-