MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
Philips received a complaint by the biomedical engineer (bme) on the v60, indicating that the ventilator was not going into standby mode.It was reported that there was no patient involvement at the time the issue was discovered.The biomedical engineer (bme) reported to the remote service engineer (rse) that the device was not going into standby mode.The rse had the bme check in service mode, and the bme confirmed that the device was showing an average air standard liter per minute (slpm) of -1.1 and -1.3.The rse then informed the bme that was the reason for device not going into standby and suggested that the bme could let the device run in normal operational mode for 2 minutes to see if the values change.The rse also suggested that the bme could also attempt to blow out the flow sensor assembly-- if the device was still failing, the rse recommended that the bme should replace the flow sensor assembly.
 
Manufacturer Narrative
H10: per good faith effort (gfe) response received 11jan2024, the biomedical engineer (bme) stated that it was not necessary to order a replacement flow sensor assembly as there was already one on hand.The bme replaced the flow sensor assembly and verified that the issue was resolved.The device passed the required performance verification tests per philips standard and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received, the complaint file will be reopened.Although requested, the defective parts were not received at this time.A supplemental report will be submitted if the part is received and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18418486
MDR Text Key331598009
Report Number2518422-2023-38489
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2024
Date Device Manufactured12/31/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-