MEDTRONIC NEUROMODULATION EXTERNAL NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_ENS_STIMULATOR |
Device Problem
High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id 3389s-40 lot# va2tk0m product type lead section d information references the main component of the system.Other relevant device(s) are: product id: 3389s-40, serial/lot #: va2tk0m, ubd: 28-feb-2026, udi#: (b)(4) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the resistance of the lead was found to be too high during the operation.A new lead was used and the surgery was completed.
|
|
Manufacturer Narrative
|
H3: the returned 3389s-40 lead was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.Analysis determined that the returned lead passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|