Updated fields: h6 (type of investigation, investigation findings, component codes and investigation conclusions).Corrected fields: h6 (health effect ¿ clinical & impact code).A getinge field service engineer (fse) investigated the customer's complaint confirmed unit will not boot up in clinical mode, when unit was turned on just has a spinning wheel.Fse was able to get into the maintenance menu but there were no faults or error codes.Replaced pcba, video generator.Reassembled the unit and turned unit on but it needed the software updated , after update unit started up as it should , ran unit thru a complete calibration and electrical safety tests and all meets factory specifications.Ran unit on test balloon with no issues,unit was returned to customer.
|