Model Number 27977 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the rad-g "turns on and then turns off." there was no patient impact or consequence reported.
|
|
Event Description
|
The customer reported the rad-g "turns on and then turns off." there was no patient impact or consequence reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.The device powered off immediately after powering on.The rad-g had a short inside the on/off button which created an electrical short across the button, resulting in the button being activated even when not physically pressed.
|
|