• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-81A
Device Problem Imprecision (1307)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2023, senseonics was made aware of an incident where a user experienced sensor inaccuracy which led to an early sensor removal.
 
Manufacturer Narrative
The customer was getting multiple out of range high glucose and sensor check alerts between 26 nov - 2 dec 2023.Raw sensor data also showed gaps and deterioration of the signal and reference channel.This is characteristic of a shorted/ disconnected led which result in a drop off in the signal/reference channels.An rma was authorized for further investigation.Upon receipt of the rma, the sensor was tested in-house, and the review of qc revealed an led crash (permanent loss of sig and ref signals), which confirms the cause of the in vivo observations.As part of resolution, the rma was authorized to offer the user a sensor replacement.G3.Date received by the manufacturer? 12 feb 2024 h3.Device evaluated by manufacturer? yes h6 method code updated to 10.H6 result code updated to 4203.H6 conclusion code updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18421928
MDR Text Key331670837
Report Number3009862700-2023-01017
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491022950
UDI-Public817491022950
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/06/2024
Device Model Number102096-81A
Device Catalogue NumberFG-7200-00-301
Device Lot NumberWP09255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-