• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TEGO¿ CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TEGO¿ CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number D1000
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2023
Event Type  malfunction  
Manufacturer Narrative
The device is available to be returned for evaluation; however, it has not yet been received.Additional contact: (b)(6).
 
Event Description
The complaint/event involved a tego¿ connector.It was reported that the plastic inside of the tego pulled out of the device and was still attached to the end of the arterial blood line upon attempting to remove the line to return the patient¿s blood.As a result the the line needed to be tugged off of the tego.The faulty tego was then replaced with a new one.There was no unexpected or prolonged care and no harm as a result of this complaint/event.
 
Manufacturer Narrative
A single used d1000 tego was returned for investigation with domed seal observed and confirmed.The top surface slit was open due to the doming which would result in leakage during use.The probable cause is a manual assembly error during manufacturing.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.D9 - date returned to mfg: 11jan2024corrected information can be found in b2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEGO¿ CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18422960
MDR Text Key331667089
Report Number9617594-2023-01216
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00840619026059
UDI-Public(01)00840619026059(17)280601(10)13684637
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K053106
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD1000
Device Lot Number13684637
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED ARTERIAL LINE, MFR UNK
-
-